Effects of space charge on the mass purification in Penning traps
Herlert, Alexander ; Borgmann, Christopher ; Fink, Daniel ; Holm Christensen, Christian ; Kowalska, Magdalena ; Naimi, Sarah
In: Hyperfine Interactions, 2011, vol. 199, no. 1-3, p. 211-220
Ajouter à la liste personnelle- Summary
- The influence of space charge on the mass selection of ions stored in a Penning trap was investigated with the ISOLTRAP experiment at CERN/ISOLDE. A mixture of 85,87Rb + ions has been used to probe the change of the experimental parameters, e.g. frequencies and amplitudes of the radiofrequency excitations, as a function of the number of ions present in the trap