In: Physical Review A, 2006, vol. 74, no. 6, p. 063415
We present an experimental study of the spectra produced by optical–radio-frequency double resonance in which resonant linearly polarized laser light is used in the optical pumping and detection processes. We show that the experimental spectra obtained for cesium are in excellent agreement with a very general theoretical model developed in our group [Weis, Bison, and Pazgalev, Phys. Rev. A 74,...
|
In: Sensors and Actuators A: Physical, 2006, vol. 129, no. 1-2, p. 1-5
Laser-pumped cesium magnetometers allow highly sensitive magnetometry at room temperature. We report on applications of that technique in biomagnetic diagnostics and in a neutron electric dipole moment (nEDM) experiment. In the biomagnetic application the magnetic field from the beating human heart is detected using a gradiometer, which reaches an intrinsic sensitivity of 80 fT/Hz1/2....
|
In: The European Physical Journal: Applied Physics, 2006, vol. 33, p. 221-224
For physical processes which express themselves as a frequency, for example magnetic field measurements using optically-pumped alkali-vapor magnetometers, the precise extraction of the frequency from the noisy signal is a classical problem. We describe herein a frequency measurement system based on an inexpensive commercially available computer sound card coupled with a software single-tone...
|
In: The European Physical Journal D, 2006, vol. 38, no. 2, p. 239-247
We discuss the design and performance of a laser-pumped cesium vapor magnetometer in the Mx configuration. The device will be employed in the control and stabilization of fluctuating magnetic fields and gradients in a new experiment searching for a permanent electric dipole moment of the neutron. We have determined the intrinsic sensitivity of the device to be 15 fT in a 1 Hz...
|
In: Journal of Research of the National Institute of Standards and Technology, 2005, vol. 110, p. 179-183
We designed laser-pumped cesium vapor magnetometers in the Mx configuration for the control and stabilization of magnetic field fluctuations and gradients in a new experiment searching for a permanent electric dipole moment of the neutron. The intrinsic sensitivity of the device was determined to be 30 fT in a measurement bandwidth of 1 Hz, limited by laser noise. In the shot noise...
|