An endoscopic detector for ultracold neutrons
Göltl, L. ; Chowdhuri, Z. ; Fertl, M. ; Gray, F. ; Henneck, R. ; Kirch, K. ; Lauss, B. ; Lefort, T. ; Mtchedlishvili, A. ; Schmidt-Wellenburg, P. ; Zsigmond, G.
In: The European Physical Journal A, 2013, vol. 49, no. 1, p. 1-9
Ajouter à la liste personnelle- Summary
- A new versatile detector for ultracold neutrons (UCN) has been built and operated which combines multi-pixel photon counters and GS10 lithium-doped scintillators. Such detectors can be very small and can be used to monitor UCN inside storage vessels or guides with negligible influence (of order 10−6 on the UCN intensity itself. We have shown that such detectors can be used in a very harsh radiation environment of up to 200Gy/h via the addition of a 4m long quartz light guide in order to place the radiation-sensitive photon counters outside the hot zone. Additionally we have measured the UCN storage times in situ in this harsh environment