Monte Carlo simulation of a mammographic test phantom

Hunt, R. A. ; Dance, D. R. ; Pachoud, M. ; Alm Carlsson, G. ; Sandborg, M. ; Ullman, G. ; Verdun, F. R.

In: Radiation Protection Dosimetry, 2005, vol. 114, no. 1-3, p. 432-435

Zum persönliche Liste hinzufügen
    Summary
    A test phantom, including a wide range of mammographic tissue equivalent materials and test details, was imaged on a digital mammographic system. In order to quantify the effect of scatter on the contrast obtained for the test details, calculations of the scatter-to-primary ratio (S/P) have been made using a Monte Carlo simulation of the digital mammographic imaging chain, grid and test phantom. The results show that the S/P values corresponding to the imaging conditions used were in the range 0.084-0.126. Calculated and measured pixel values in different regions of the image were compared as a validation of the model and showed excellent agreement. The results indicate the potential of Monte Carlo methods in the image quality-patient dose process optimisation, especially in the assessment of imaging conditions not available on standard mammographic units