In: Nuclear Instruments and Methods in Physics Research Section A, 2009///doi: 10.1016/j.nima.2009.07.047
The existence of a mirror world holding a copy of our ordinary particle spectrum could lead to oscillations between the neutron (n) and its mirror partner (n′). Such oscillations could manifest themselves in storage experiments with ultracold neutrons whose storage lifetime would depend on the applied magnetic field. Here, extended details and measurements from the first dedicated experimental...
|
In: Physics Letters B, 2015, vol. 745, p. 58–63
|
In: The European Physical Journal D, 2015, vol. 69, no. 8, p. 1-12
|
In: The European Physical Journal D, 2015, vol. 69, no. 8, p. 1–12
We report on the design and performance of a highly sensitive combined ³He/Cs magnetometer for the absolute measurement of magnetic fields. The magnetometer relies on the magnetometric detection of the free spin precession of nuclear spin polarized ³He gas by optically pumped cesium magnetometers. We plan to deploy this type of combined magnetometer in an experiment searching for a permanent...
|
In: Journal of Research of the National Institute of Standards and Technology, 2005, vol. 110, p. 179-183
We designed laser-pumped cesium vapor magnetometers in the Mx configuration for the control and stabilization of magnetic field fluctuations and gradients in a new experiment searching for a permanent electric dipole moment of the neutron. The intrinsic sensitivity of the device was determined to be 30 fT in a measurement bandwidth of 1 Hz, limited by laser noise. In the shot noise...
|
In: The European Physical Journal A, 2015, vol. 51, no. 11, p. 1–10
We report on the design and first tests of a device allowing for measurement of ultracold neutrons polarisation by means of the simultaneous analysis of the two spin components. The device was developed in the framework of the neutron electric dipole moment experiment at the Paul Scherrer Institute. Individual parts and the entire newly built system have been characterised with ultracold...
|
In: Journal of Applied Physics, 2014, vol. 116, no. 8, p. 084510
The Surrounding Field Compensation (SFC) system described in this work is installed around the four-layer Mu-metal magnetic shield of the neutron electric dipole moment spectrometer located at the Paul Scherrer Institute. The SFC system reduces the DC component of the external magnetic field by a factor of about 20. Within a control volume of approximately 2.5 m × 2.5 m × 3 m,...
|
In: Physical Review A, 2006, vol. 74, no. 6, p. 063415
We present an experimental study of the spectra produced by optical–radio-frequency double resonance in which resonant linearly polarized laser light is used in the optical pumping and detection processes. We show that the experimental spectra obtained for cesium are in excellent agreement with a very general theoretical model developed in our group [Weis, Bison, and Pazgalev, Phys. Rev. A 74,...
|
In: The European Physical Journal D, 2006, vol. 38, no. 2, p. 239-247
We discuss the design and performance of a laser-pumped cesium vapor magnetometer in the Mx configuration. The device will be employed in the control and stabilization of fluctuating magnetic fields and gradients in a new experiment searching for a permanent electric dipole moment of the neutron. We have determined the intrinsic sensitivity of the device to be 15 fT in a 1 Hz...
|
In: The European Physical Journal D, 2015, vol. 69, no. 11, p. 1–11
We report on extensive studies on the intrinsic sensitivity of a combined ³He/Cs magnetometer. The magnetometer relies on the detection of the free spin precession of nuclear spin polarized ³He by optically pumped cesium magnetometers. We characterize the relevant processes involved in the detection and quantify their impact on the total sensitivity of the magnetometer. An expression is derived...
|