In: The European Physical Journal A, 2015, vol. 51, no. 11, p. 1-10
|
In: Physical Review A, 2019, vol. 99, no. 4, p. 042112
Magnetic-field uniformity is of the utmost importance in experiments to measure the electric dipole moment of the neutron. A general parametrization of the magnetic field in terms of harmonic polynomial modes is proposed, going beyond the linear- gradients approximation. We review the main undesirable effects of nonuniformities: depolarization of ultracold neutrons and Larmor frequency shifts...
|
In: Applied Physics B, 2014, vol. 115, no. 2, p. 257-262
|
In: The European Physical Journal A, 2013, vol. 49, no. 1, p. 1-9
|
In: The European Physical Journal A, 2015, vol. 51, no. 11, p. 1–10
We report on the design and first tests of a device allowing for measurement of ultracold neutrons polarisation by means of the simultaneous analysis of the two spin components. The device was developed in the framework of the neutron electric dipole moment experiment at the Paul Scherrer Institute. Individual parts and the entire newly built system have been characterised with ultracold...
|
In: Physical Review Letters, 2015, vol. 115, no. 16, p. 162502
We describe a spin-echo method for ultracold neutrons (UCNs) confined in a precession chamber and exposed to a |B0|=1 μT magnetic field. We have demonstrated that the analysis of UCN spin-echo resonance signals in combination with knowledge of the ambient magnetic field provides an excellent method by which to reconstruct the energy spectrum of a confined ensemble of neutrons. The method...
|
In: Physics Letters B, 2014, vol. 739, p. 128–132
The neutron gyromagnetic ratio has been measured relative to that of the 199Hg atom with an uncertainty of 0.8 ppm. We employed an apparatus where ultracold neutrons and mercury atoms are stored in the same volume and report the result γn/γHg=3.8424574(30)γn/γHg=3.8424574(30).
|
In: Journal of Applied Physics, 2014, vol. 116, no. 8, p. 084510
The Surrounding Field Compensation (SFC) system described in this work is installed around the four-layer Mu-metal magnetic shield of the neutron electric dipole moment spectrometer located at the Paul Scherrer Institute. The SFC system reduces the DC component of the external magnetic field by a factor of about 20. Within a control volume of approximately 2.5 m × 2.5 m × 3 m,...
|
In: Physics Procedia, 2011, vol. 17, p. 159-167
The measurement of the neutron electric dipole moment (nEDM) constrains the contribution of CP-violating terms within both the Standard Model and its extensions. The experiment uses ultracold neutrons (UCN) stored in vacuum at room temperature. This technique provided the last (and best) limit by the RAL/Sussex/ILL collaboration in 2006: dn < 2:9 × 10⁻²⁶ e cm (90% C.L.). We aim to improve...
|
In: Physics Procedia, 2011, vol. 17, p. 259-267
The nEDM experiment hosted at the Paul Scherrer Institute is the flagship project at the new ultracold neutron facility. Estimations of systematic effects for the determination of the neutron electric dipole moment play an important role in this project. Experimental studies are supported by Monte Carlo simulations using the MCUCN code. Here we briefly present first results on the experimental...
|